The alarm goes off like an air raid siren – everybody starts to panic – somebody starts to do the saturation countdown. There is nothing quite as distressing for the anesthesiologist or intensivist than for the ventilator to pressure cycle and fail to deliver tidal volumes due to high airway pressure.
Generally high pressures are caused by one of three things – a problem with the equipment (kinked tubing, patient biting the tubing etc.), an airway resistance problem (e.g. bronchospasm) or a pulmonary compliance problem (e.g. consolidation or pulmonary edema) or a combination of these. The first thing that the clinician should do when there pressure alarm goes off – is to silence the alarm and increase the Pmax.
Then go looking for the problem: start at the mouth and work your way back to the machine. If you can’t find a fault, put the patient on a manual breathing circuit and commence ventilation. If the patient is easy to bag, there is a machine problem, if difficult – then there is a problem with pulmonary resistance or compliance. In this first tutorial I look at assessing airway resistance. I do this in two ways. First I discuss peak to plateau pressure gradients and then look at airway resistance: dynamic versus static and how to calculate it. I will finish the discussion in the next tutorial.